Beschreibung

PRESSURE TRANSMITTER UNIVERSAL CA for diaphragm seal operation, hygienic see data sheet D4-018-12 pressure sensor: piezoresistive for gauge pressure measuring range -1...9 bar, overload limit 100 bar measuring accuracy: linearity, hysteresis, repeatability <= 0.2 % of span output signal: 4...20 mA, 2-wire technology electrical connection: circular connector M12x1 (4 pin) design: with temperature decoupler process connection: Tri-Clamp, DN 1 1/2" / PN 25 material stainless steel mat.-no. 1.4404/1.4435 (316L) system filling: Synthetic oil, free of silicon FD1, FDA listed hygienic design per EHEDG Doc.8 and ASME BPE SF3: surface roughness (wetted parts) - turned parts Ra <= 0.76 μm - diaphragm Ra <= 0.38 μm - weld seam Ra <= 0.76 μm process temperature: -20...+160°C Components marked acc. to material inspection certificate per EN 10204-3.1, wetted parts TEST REPORT per DIN EN 10204 2.2 CALIBRATION PROTOCOL: with 5 measuring points